PDL 1100 Series Specifications |
|||
Gas Service |
Del. Press. |
Flow Rate (Air) |
Liquid Cyl. To use |
Oxygen |
85 psig |
Cyl. Dependent |
All liquid cylinders can be used with the PDL1100 series. |
Nitrogen |
|||
Argon |
|||
CO2 without a Heater |
35 cfh |
||
CO2 with a Heater |
Cyl. Dependent |
||
PDL 2100 Series Specifications |
||||
Del. Press |
Flow Rate (Air) |
Liquid Cyl. To use |
||
200 psig |
Cyl. Dependent |
To obtain maximum delivery pressure, liquid cylinders with 350 psig or 500 psig relief valve settings required |
||
35 cfh |
||||
Cyl. Dependent |
||||
How To Order: Specify Control Type, Number of Cylinders and Gas Service |
|||
| Example 1: PDL1102-540 represents PDL with two (2) cylinders for oxygen gas service (5-85PSIG) | |||
| Example 2: PDL2102-540 represents PDL with two (2) cylinders for oxygen gas service (10-200 PSIG) | |||
Control Type |
Cyl Series - No. Of Cylinders |
Identifier |
Gas Service |
PDL 1100 |
1102 (2 Cyl.) |
580A |
Argon |
PDL 2100 |
2102 (2 Cyl.) |
||
Gas withdrawal rate table from standard liquid cylinders |
|||
Gas Service |
Continuous Withdrawal Rate |
Gas Service |
Continuous Withdrawal Rate |
Oxygen Nitrogen Argon |
1 cylinder per side: 2 cylinders per side: 3 cylinders per side: |
Carbon Dioxide |
1 cylinder per side 110 cfh 2 cylinders per side 150-160 cfh 3 cylinders per side 200-220 cfh |
Visual and Audible Alarms for the PDL1100 Series |
|
Model |
Gas Service |
RW-12VS-117 |
Oxygen |
| Argon | |
| Nitrogen | |
| Carbon Dioxide | |
Visual and Audible Alarms for the PDL2100 Series |
|
Model |
Gas Service |
RW-12VS-225 |
Oxygen |
| Argon | |
| Nitrogen | |
| Carbon Dioxide | |